Cookie Img
Home » Products » Earthquake Monitoring Instruments » Broad Band Seismometer

Broad Band Seismometer

Broad Band Seismometer
Broad Band Seismometer
Product Code : 151-120
Product Description

Model 151-30 / 151-60 /151-120



Features:

  • High Performance Broadband Seismometer

  • Low Self-Noise Over a Wide Bandwidth

  • Portable Design

  • Low Power Consumption

  • Force-Balance Electronic Feedback

  • Robust Mass Locking Mechanism

  • Easy Installation

Applications:

  • Local and Regional

  • Broadband Surface or Posthole Installation

  • GSN Installation



The 151 series Broadband Seismometer, “Observer”, is a force-balance feedback sensor available with frequency bandwidth 0.0083 Hz (120 sec) - 50 Hz, and 0.016 Hz (60 sec) - 50 Hz, flat to velocity. The “Observer” with with frequency bandwidth 0.033 Hz (30 sec) - 50 Hz is available upon request. The 151 “Observer” contains three independent sensors (one vertical and two horizontal) with built-in electronic feedback circuit, control and power conversion circuits, featuring low noise, large dynamic range and easy installation and use.

The Observer has built-in leveling and mass lock/unlock facilities. The leveling mechanism includes two bubble levels, three adjustable feet and three locknuts which are located on the seismometer’s chassis (see photo below).

The Observer has built-in mass zero-position adjusting mechanism to perform automatic mass adjustment. As soon as the seismometer is powered it checks the zero position of each component’s mass and automatically adjusts the zero position if needed. Monitoring and adjustment of the mass can also be performed via the 130 family High Resolution Recorders using Sensor Control Board, RT527-B01.

The Observer is a low noise seismometer (see below Power Spectral Density plot). The low self noise performance makes the 151-120 an ideal seismometer for local, regional, and global seismicity studies in different installation configurations, including observatory and portable, surface and posthole applications.


Noise Model for REF TEK 151



Sensor Type

Triaxial, orthogonal

Feedback

Force balance with capacitive displacement transducer

Frequency Response

0.0083 Hz (120 sec) - 50 Hz (see diagram in manual), Model 151-120
0.0166 Hz (60 sec) - 50 Hz, Model 151-60
0.033 Hz (30 sec) - 50 Hz, Model 151-30 (optional upon request)

Sensitivity

2000 V/m/s

Output Signal

+/- 20 V peak-to-peak differential

Dynamic Range

>140 dB @ 5Hz

Output Impedance

<100 Ohms

Calibration

Coil resistance 1,000 Ohms, sensitivity: 10 m/s2/A

Leveling

Two Integrated bubble levels. Three adjustable leveling feet with locknuts

Mass Position

Remote monitoring and adjustable zero-point using 3 independent voltage outputs

Mass Locking

Mechanical lock/unlock on outside of the case

Distortion

Total distortion less than -80 dB

Cross Axis Coupling

< 1%

Mechanical Zero

No need for adjustment within ambient temperature fluctuation +/- 20o C

Power-Fail Protection

Built-in

Signal Overload Protection

Built-in

Auto-recovery Time

Full operation recovery within 10 minutes after either power fail or signal overload

Supply Voltage

9 to 18 VDC

Operating Temperature

-20o C to +60o C

Power Consumption

~ 1.1 W

Water Tight Integrity

IP67 for outdoor use and immersion resistance

Case

Aluminum, Powder Coated

Dimensions

Diameter: 9.45 " (24 cm); Height: 10.6 " (27 cm)

Weight

26.4 lbs (12 kg)

Connector

PTO2A 14-19S

Alignment

Removable orientation North(brass) / South(silver) pins screw into the base

Accessories (optional)

Rugged transportable hardigg case. Flexible insulation bag (P/N 151-9800)


Standard Observer Kit

The standard Observer kit includes: a) Seismometer; b) T-type Allen wrench key for leveling; c) Aluminum carrying case with pre-cut foam; d) orientation pins




Optional Ruggedized Transport Case

The Observer is available with an optional ruggedized transportation case and thermal insulation bag.

The Observer's transportation case is a ruggedized, durable, lockable case with handles and removable casters, designed for multiple shipments by any means of transportation in a harsh environment.

The transportation case has pre-cut foam to house the 151 Observer with carrying case, removable casters and accessories, including sensor cable and thermal insulation bag.

The Observer thermal insulation bag is fabricated from Microfleece material with primaloft insulation, providing protection from wind and daily temperature fluctuations in the vault to insure the highest quality data.




TOSHNIWAL TECHNOLOGIES PVT. LTD.

Contact Us

( *represents compulsory fields )

Tips on getting accurate quotes. Please include product name, order quantity, usage, special requests if any in your inquiry.



TOSHNIWAL TECHNOLOGIES PVT. LTD. All Rights Reserved.